Interference screws manufactured from magnesium display similar primary stability for soft tissue anterior cruciate ligament graft fixation compared to a biocomposite material - a biomechanical study

Deichsel, Adrian; Glasbrenner, Johannes; Raschke, Michael J.; Klimek, Matthias; Peez, Christian; Briese, Thorben; Herbst, Elmar; Kittl, Christoph

Research article (journal) | Peer reviewed

Abstract

PURPOSE - METHODS - RESULTS - CONCLUSION; Biodegradable interference screws (IFS) can be manufactured from different biomaterials. Magnesium was previously shown to possess osteoinductive properties, making it a promising material to promote graft-bone healing in anterior cruciate ligament reconstruction (ACLR). The purpose of this study was to compare IFS made from magnesium to a contemporary biocomposite IFS.; In a porcine model of ACL reconstruction, deep porcine flexor tendons were trimmed to a diameter of 8 mm, sutured in Krackow technique, and fixed with either 8 × 30 mm biocomposite IFS (Bc-IFS) or 8 × 30 mm magnesium IFS (Mg-IFS) in an 8 mm diameter bone tunnel in porcine tibiae. Cyclic loading for 1000 cycles from 0 to 250 N was applied, followed by load to failure testing. Elongation, load to failure and stiffness of the tested constructs was determined.; After 1000 cycles at 250 N, elongation was 4.8 mm ± 1.5 in the Bc-IFS group, and 4.9 mm ± 1.5 in the Mg-IFS group. Load to failure was 649.5 N ± 174.3 in the Bc-IFS group, and 683.8 N ± 116.5 in the Mg-IFS group. Stiffness was 125.3 N/mm ± 21.9 in the Bc-IFS group, and 122.5 N/mm ± 20.3 in the Mg-IFS group. No significant differences regarding elongation, load to failure and stiffness between Bc-IFS and Mg-IFS were observed.; Magnesium IFS show comparable biomechanical primary stability in comparison to biocomposite IFS and may therefore be an alternative to contemporary biodegradable IFS.

Details about the publication

JournalJournal of Experimental Orthopaedics
Volume10
Article number103
StatusPublished
Release year2023 (10/10/2023)
Language in which the publication is writtenEnglish
DOI10.1186/s40634-023-00663-3
Keywords Interference screw; Biocomposite; Magnesium; Acl reconstruction; Biomechanics

Authors from the University of Münster

Briese, Thorben
Clinic for Accident, Hand- and Reconstructive Surgery
Deichsel, Adrian
Clinic for Accident, Hand- and Reconstructive Surgery
Glasbrenner, Johannes
Clinic for Accident, Hand- and Reconstructive Surgery
Herbst, Elmar
Clinic for Accident, Hand- and Reconstructive Surgery
Kittl, Christoph
Clinic for Accident, Hand- and Reconstructive Surgery
Klimek, Matthias
Clinic for Accident, Hand- and Reconstructive Surgery
Peez, Christian
Clinic for Accident, Hand- and Reconstructive Surgery
Raschke, Michael Johannes
Clinic for Accident, Hand- and Reconstructive Surgery