False labelling due to quenching failure of N-hydroxy-succinimide-ester-coupled dyes.

Wang W, Ackermann D, Mehlich AM, König S

Forschungsartikel (Zeitschrift) | Peer reviewed

Zusammenfassung

In comparative fluorescence gel electrophoresis experiments, cross-talk was detected. It was traced back to a failure in the quenching process in typical labelling protocols. Despite a huge excess of potential reaction sites for the N-hydroxy-succinimide-ester-coupled dye, sufficient active dye molecules were available after the quenching step to label protein molecules un-specifically. It could be shown that only a 100-fold increase in the amount of quencher will silence residual dye to such an extent that no artificial signals are detected.

Details zur Publikation

FachzeitschriftProteomics
Jahrgang / Bandnr. / Volume10
Ausgabe / Heftnr. / Issue7
Seitenbereich1525-1529
StatusVeröffentlicht
Veröffentlichungsjahr2010
Sprache, in der die Publikation verfasst istEnglisch
DOI10.1002/pmic.200900679
StichwörterProteomics; Fluorescent Dyes; Carbocyanines; Esters; Succinimides; Electrophoresis Gel Two-Dimensional; Models Molecular; Proteomics; Fluorescent Dyes; Carbocyanines; Esters; Succinimides; Electrophoresis Gel Two-Dimensional; Models Molecular