Ablation of symptomatic uterine fibroids with the Mirabilis system for rapid noninvasive ultrasound-guided high-intensity focused ultrasound (HIFU): a prospective observational clinical studyOpen Access

Tonguc, T; Savchenko, O; Ramig, O; Stader, JM; Kießling, F; Küppers, J; Egger, EK; Thudium, M; Martin, P; Poll, W; Schild, HH; Conrad, R; Essler, M; Mustea, A; Strunk, HM; Marinova, M.

Forschungsartikel (Zeitschrift) | Peer reviewed

Zusammenfassung

Objectives: Uterine fibroids often lead to symptoms that negatively impact health-related quality of life (HRQOL). High-intensity focused ultrasound (HIFU) has emerged as a promising noninvasive treatment for reducing fibroid size and symptoms. The Mirabilis system for ultrasound (US)-guided HIFU introduces a novel technique known as 'shell ablation'. This study evaluates the feasibility and efficacy of Mirabilis in a clinical setting, focusing on clinical outcomes. Materials and methods: Sixteen patients with 23 uterine fibroids were treated with the Mirabilis system. Follow-up assessments included US and MRI at baseline, 6 weeks, 3, 6 and 9 months, and 1 year after HIFU. Changes in symptoms and QOL were evaluated using the Uterine Fibroid Symptom and HRQOL Questionnaire. Results: A significant reduction in fibroid volume was observed after HIFU (baseline 182.1 ± 49.3 ml; 1 year: 76.0 ± 37.9 ml, p < 0.001). The symptom severity score significantly declined (baseline 57.2 ± 3.8; 1 year: 30.2 ± 4.9, p < 0.001), correlating with a significant improvement in HRQOL (baseline 47.0 ± 3.9, 1 year: 71.8 ± 5.3, p < 0.001). Conclusion: HIFU with the portable Mirabilis system is a feasible and safe noninvasive treatment for symptomatic uterine fibroids in an outpatient setting. This approach allows efficient and rapid ablation even for large fibroids, significantly reducing fibroid volume and symptoms.

Details zur Publikation

FachzeitschriftRadiologia Medica
Jahrgang / Bandnr. / Volume130
Ausgabe / Heftnr. / Issue5
Seitenbereich629-637
StatusVeröffentlicht
Veröffentlichungsjahr2025
Sprache, in der die Publikation verfasst istEnglisch
DOI10.1007/s11547-025-01972-6
Link zum Volltexthttps://link.springer.com/article/10.1007/s11547-025-01972-6
StichwörterFibroid-associated symptoms; High-intensity focused ultrasound; Mirabilis system; Mobile device; Shell ablation.

Autor*innen der Universität Münster

Conrad, Rupert
Klinik für Psychosomatische Medizin und Psychotherapie